CHECKING AND REPLACING AN ELECTROCHEMICAL OXYGEN SENSOR
25 - 5
ETC00781(4) Series 100 e 02/2004
b) IR / Oxygen Measurement combined
P Open the analyzer housing (see chapter 23.).
P Disconnect the connector for the sensor from “P2” of circuit board “OXS” (see Fig. 25-5).
P Unscrew both fastening screws for the fitting (see Fig. 25-4) and remove the fitting
completely with the sensor.
Fig. 25-4: Inside View BINOS
®
100 2M
(combined IR / Oxygen Measurement, assembly similar in BINOS
®
100 M)
REPLACING THE SENSOR
Fastening
screws
Fitting
Gas connections
Front panel