8 - 12
ENTRY OF SYSTEM PARAMETERS
ETC00265(4) Series 100 e 02/2004
8.14 Reset
The reset operation restores the settings of the analyzer to the parameters and calibration factors
set in our factory at the time of its manufacture.
This is equivalent to switching off the electrical supply line and switching off the battery buffering
of the RAM’s by removing the battery jumper, J7.
All parameters and calibration factors entered by the user will be lost whenever a reset
operation is performed.
The currently valid user identification code must be entered before a reset will be executed; this
will prevent inadvertent resets.
Entry is performed using
followed by
Whenever a reset operation is initiated, the analyzer operating program will be restarted, just as
it is when the instrument is first switched on (see chapter 6).
Jumper J6, which activates the watchdog circuitry must be inserted if the
reset operation is to be correctly executed.
RESET