8 - 4
SETTING SYSTEM PARAMETERS
ETC00781(4) Series 100 e 02/2004
8.4 Hold
The analyzer function HOLD is used to lock the analog signal outputs and the concentration
limits during a calibration procedure to hold the last measured values.
Input value 0: The outputs remain unlocked.
Input value 1: The outputs will be locked.
Change values using
and
8.5 Automatic Calibration
For operation with optional, internal or external solenoid valves it can be selected, if there a time-
controlled (automatic) resp. a remote-controlled calibration is possible or not (in combination with
solenoid valves controlled via digital outputs).
Input value 0: Time-controlled / remote-controlled calibration is not possible
Input value 1: Time-controlled / remote-controlled calibration is possible
Change values using
and
HOLD / AUTOMATIC CALIBRATION