8 - 8
SETTING SYSTEM PARAMETERS
ETC00781(4) Series 100 e 02/2004
8.9 Flushing Period
For calibration, the gas paths must be supplied with sufficient calibration gas. The flushing period
has to be fixed adequate; perform calibration only after a suitable flushing period (the calibration
gas flow should be identical with sample gas flow).
This period may be set within the range 0 - 99 sec. depending on calibration conditions.
Change values using
and
8.10 User Code
Factory default setting is "1".
To prevent parameter alterations by unauthorized persons, the operator may specify another
password (user code).
Set the code using
and
Please take care for filing the user code.
FLUSHING PERIOD / USER CODE